Ducted Fume Hood

50" Wide Ducted Fume Hood

Model # SS-350-E

Loading the player...

Durable & Reliable Construction . The 50” Wide Ducted Fume Hood, commonly referred to as an exterior-venting exhaust hood, is constructed of durable materials that are resistant to many solvents and chemicals. These qualities make it a suitable fume control solution in a variety of industries including manufacturing, laboratories, pharmacies, research and development facilities, and medical environmentsThis configuration is frequently used when processes involve volatile organics, solvent vapors, epoxy fumes, powder and capsule filling, acid etching, and a variety of additional applications for which filtration media may not be an appropriate strategy.

This benchtop Ducted Fume Hood fits on a standard 24” deep countertop; therefore, no specialized stands or custom work surfaces are typically required.

Customization to Your Needs. Sentry Air Systems Engineering Team can design customer-specific hoods that address unique applications. You can see some of our customized hoods here. Options include, but are not limited to:

Material Options (e.g. metal, esd-safe, PVC)
Size Options (width, height, depth, access area dimensions)
Custom Cutouts (e.g. sink, machinery, microscopes)
Dual-Entry Configuration
Custom Illumination Options (e.g. Black Light)
Accessories for Hanging Equipment

This unit includes a 8” Exhaust Collar and does not include ducting or fan. Ask Sentry Air Systems Applications Specialists for assistance in selecting a suitable fan. Crating for shipping is available as well. To view fume hoods that include a fan and filtration media, visit our Ductless Fume Hood page.

Specifications
Hood Dimensions:
O.D: 24"D x 52W" x 25"H
I.D.: 23.5"D x 50"W x 24"H
Hood Material:
Sides: .250" White Acrylic
Base: .500" White High Density Polyethylene
Top: .500" Clear Polycarbonate
Front: .250" Clear Polycarbonate
Total Weight:
Approx. 75 lbs.
Exhaust Connection:
8" OD Outlet Collar (ducting and fan excluded)
Warranty:
Limited two-year warranty from date of shipment on defects due to materials or workmanship.
Max Operating Temperature:
Not to exceed 170 degrees Fahrenheit

All specifications are nominal and subject to change without notice. Not intended for explosive or highly flammable applications. (i.e. when the Lower Flammability Limit is exceeded for any given chemical). Please consult the specific MSDS for questions regarding flammability limits. The end user is ultimately responsible for conforming to all applicable exhaust ventilation regulations. Consult local, state and federal agencies regarding the suitability of your application prior to venting fumes to the environment. It is your ultimate responsibility to ensure the safety and suitability of your application. For more information, contact a local safety professional (certified industrial hygienist) for conformance and suitability.